    7400安規(guī)綜合測試儀器

    • 測溫范圍:
    1. 詳細(xì)信息
    2. 選項(xiàng)卡二
    3. 選項(xiàng)卡三
    4. 選項(xiàng)卡四
    5. 選項(xiàng)卡五

     

    SPECIFICATIONS
    MODEL
    7450
    7451
    7452
    AC WITHSTAND VOLTAGE
    Output Rating 5KVAC/100mA Constant 500VA
    Range
    Resolution
    Accuracy
    Output Voltage, ACV
    0 - 5000
    1
    ±(2% of setting + 5V)
    Output Frequency 50Hz/60Hz ±100ppm, User Selection
    Output Waveform Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5
    Output Regulation ± (1% of output + 5V), From no load to full load
    SETTINGS
    Hi-Limit AC Current, mA
    0 - 99.99
    0.01
    ±(2% of setting + 3counts)
    Lo-Limit AC Current, mA
    0 - 9.999
    0.001
    ±(2% of setting + 3counts)
    Ramp Time, second
    0.1 - 999.9
    0.1
    ±(0.1% + 0.05sec)
    Dwell Time, second
    0, 0.3 - 999.9
    (0=continuous)
    0.1
    Arc Detection 1 - 9 ranges (9 is the most sensitivity)
    AC Current Offset, mA 0 - 2.000 or Auto Set
    DC WITHSTAND VOLTAGE (7451,7452 ONLY)
    Output Rating 6KVDC/9999µA
    Output Voltage, DCV 0 - 6000 1 ±(2% of setting + 5V)
    Output Ripple <5%(6KV/9999µA at Resistive Load)
    SETTINGS
    Hi-Limit DC Current, µA
    0 - 9999
    1
    ±(2% of setting + 3counts)
    Lo-Limit DC Current, µA
    0 - 999.9
    0.1
    ±(2% of setting + 3counts)
    Ramp Time, second
    0.1 - 999.9
    0.1
    ±(0.1% + 0.05sec)
    Dwell Time, second
    0, 0.4 - 999.9
    (0=continuous)
    0.1
    Arc Detection 1 - 9 ranges (9 is the most sensitivity)
    DC Current Offset, µA 0 - 200.0 or Auto Set
    Ramp High Current ON/OFF, User Selection
    Charge - Low Current, µA 0 - 350.0 or Auto Set
    Discharge Time < 200 msec
    INSULATION RESISTANCE
    Output Rating 1KVDC/9999MΩ
    Output Voltage, DCV
    100 - 1000
    1
    ±(2% of setting + 3V)
    Hi-Limit Resistance, MΩ
    0, 1 - 9999 (0=OFF)
    1
    ±(2% of setting+2counts)at 500-1000VDC, 1 -1000 MΩ
    ±(8% of setting+2counts)at 500-1000VDC, 1000 -9999 MΩ
    ±(8% of setting+2counts)at 100-500VDC, 1 -1000 MΩ
    Lo-Limit Resistance, MΩ
    1 - 9999
    1
    Delay Time, second
    0, 0.5 - 999.9
    (0=continuous)
    0.1
    ±(0.1% + 0.05sec)
    Charge - Low Current, ?A
    0 - 3.500 or Auto Set
    GROUND BOND (7452 ONLY)
    Output Rating 30A/600mΩ, 8V
    Output AC Current, A
    3.00 - 30.00
    0.01
    ± (2% of setting +2 counts)
    Output AC Voltage, V
    3.00 - 8.00
    0.01
    ± (2% of setting +3 counts)
    Output Frequency 50Hz/60Hz ±100ppm, User Selection
    SETTINGS
    Hi-Limit Resistance, mΩ
    0 - 150 (3 - 30A)
    1
    ± (2% of setting +2 counts)
    Lo-Limit Resistance, mΩ
    0 - 600 (3 - 10A)
    Lead Resistance Offset, mΩ
    0 -200
    1
    ± (2% of setting +2 counts)
    Dwell Time, second
    0, 0.5 - 999.9 sec (0 = continuous)
    0.1
    ±(0.1% + 0.05sec)
    MEASUREMENT
    AC/DC Voltage, KV
    0 - 6.00
    0.01
    ±(1.5% of reading + 1count)
    DC Voltage, V (IR)
    100 - 1000
    1
    ±(1.5% of reading + 3count)
    AC Current Range, mA
    0-3.500
    0.001
    ±(2% of reading + 3counts)
    3.00-99.99
    0.01
    ±(2% of reading + 6counts)
    DC Current Range, µA
    0-350.0
    0.1
    ± (2% of reading + 3counts)
    300 - 3500
    1
    ± (2% of reading + 2counts)
    3000-9999
    10
    ± (2% of reading + 6counts)
    AC Current, A (GB)
    0 - 35.00
    0.01
    ±(2% of reading+2counts)
    Resistance, , MΩ(IR)
    1 - 9999
    (Auto Range)
    ±(2% of reading+2counts)at 500-1000VDC, 1 -1000 MΩ
    ±(8% of reading+2counts)at 500-1000VDC, 1000 -9999 MΩ
    ±(8% of reading+2counts)at 100-500VDC, 1 -1000 MΩ
    Resistance, , mΩ(GB)
    0 - 600
    1
    ±(2% of reading+2counts)
    GENERAL
    Input Voltage AC 100/115/200/230Vac±15%, 50/60Hz ± 5%, Fuse 15A
    PLC Remote Control Input : Test , Reset , Memory1.2.3
    Output : Pass , Fail , Processing
    Memory 50 memories, 8 steps/memory
    Display 20×2 LCD with back light
    Key Lock To prevent unauthorized alteration of the test Parameters
    Calibration Build-in software and external calibrated meters
    Alarm Volume Setting Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume.
    Rear Panel Output
    YES
    YES
    YES
    Built-in Scanner
    -
    -
    -
    External Scanner
    -
    YES
    YES
    Smart GFI
    YES
    YES
    YES
    Analog Meter
    YES
    YES
    YES
    RS232 Interface
    -
    Option
    Option
    GPIB Interface
    -
    YES
    YES
    Printer Interface
    -
    Option
    Option
    TUV-GS listed
    YES
    -
    YES
    CE Compliance
    YES
    -
    YES
    Environment
    0-40℃, 20-80%RH
    Dimension (W×H×D)
    430×133×400 (mm)
    430×133×400 (mm)
    430×133×500 (mm)
    Net Weight
    28Kg
    28Kg
    32Kg
    STANDARD ACCESSORIES
    Power Cord(15A)
    ×1
    Fuses
    ×2 (Including a spare contained in the fuse holder)
    Hipot Test Lead ,1.5m(1101)
    ×1
    ×1
    ×1
    Hipot Return Lead ,1.8m(1102)
    ×1
    ×1
    -
    Ground Bond Test Lead 30A
    ,1.5m(1103)
    -
    -
    ×1
    Ground Bond Return Lead 30A
    ,1.5m(1104)
    -
    -
    ×1
    *product specifications are subject to change without notice.